1、 Application scope:
The Multi gas corrosion test chamber is widely used in enterprises and institutions such as aerospace, automobiles and motorcycles, military equipment and weapons, electrical and electronics, chemical metallurgy, quality inspection and measurement, postal and telecommunications, ships and railways, as well as universities, colleges and research institutions. 2、 Execution standards
1. The reference standard for mixed gas testing equipment is GB/T 2423.51-2012 GB/T 2423.19-2013、GB/T 2423.20-2014、GB/T19745-2005;
2. Design and manufacture in accordance with QC/T417.1-2001 4.24 Flow Gas Corrosion Test requirements;
3. Suitable for individual or mixed gas testing of H2S, SO2, NO2, and CL2.
3、 Technical parameters:
Explanation of Climate Box Parameters
Inner container volume size ≈ 266 L, width 630mm X depth 630mm X height 670mm
The external dimensions are approximately 800mm wide, 1500mm deep, and 1900mm High;
(Note: The external dimensions are for reference only, and the actual 3D drawings in the technical agreement shall prevail)
The installed power is about 9.5kw
Rated current of approximately 14.71A
Power supply conditions 3 ¢ 380VAC ± 10%, 50Hz three-phase four wire+protective grounding
Voltage allowable fluctuation range: 10%
Allowable frequency fluctuation range: (50 ± 0.5) Hz
The grounding resistance of the protective grounding wire is less than 4 Ω
Require users to configure corresponding capacity air or power switches for the equipment at the installation site, and this switch must independently control the use of the equipment
Cooling method: air cooling
Temperature range 10 ℃~+90 ℃
The humidity range is 10% r.h. to 98% r.h
Temperature accuracy fluctuation: ± 0.5K; Uniformity: ± 2K; Temperature uniformity in the experimental area: ± 2.0K
Humidity accuracy humidity fluctuation: ± 3.0 r.h. to ± 5.0% r.h
Dew point range+5.0 ℃~+89 ℃
Heating and cooling rate, temperature change rate, heating up 0.5K/min, cooling down 0.2K/min
Note: Kelvin temperature: thermodynamic absolute temperature scale, unit K (Kelvin), 0 K is absolute 0 degrees.
Kelvin (K) is one of the seven fundamental units in the International System of Units
Celsius temperature: The point of coexistence of solid, liquid, and gas phases of pure water at one standard atmospheric pressure (273.15K) is taken as 0 degrees, and the boiling point of pure water at one standard atmospheric pressure is 100 degrees. The difference between these two quantities, 1/100, is taken as 1 degree.
The relationship between Kelvin and Celsius: 1K increments in Kelvin are numerically equal to 1 ℃ increments in Celsius.
Explanation of Corrosion Test Parameters
Corrosive gases are suitable for individual or mixed gas testing of H2S, SO2, NO2, and CL2
The maximum concentration of H2S, SO2, and NO2 in the gas is 25ppm, and the concentration of Cl2 is 1ppm
The minimum concentration of gas is determined by the background concentration of the ambient air
Gas distribution is regulated by a thermal mass flow controller with a flow rate of 0.2-10 ml/min, with a standard accuracy of ± 0.8% Rd plus ± 0.2% FS
Temperature range 15 ℃~+85 ℃
Humidity range: 15% r.h. to 90% r.h
Temperature accuracy fluctuation: ± 0.5K; Uniformity: ± 2K; Temperature uniformity in the experimental area: ± 2.0K
Humidity accuracy humidity fluctuation: ± 3.0 r.h. to ± 5.0% r.h; Real time humidity deviation at the center of the experimental area: maximum ± 5%
Dew point range+10 ° C to+37 ° C
The ventilation system includes a suction device with a monitor, control valve, and flow meter. The adjustable air exchange rate per hour can be adjusted between 1-6 times the volume of the internal container