日韩无码国产精品,91精品国产福利姬喷水福利在线,色噜噜日韩精品欧美一区二区,亚洲精品中文字幕无码久久久久久

Welcome to Tianjin Guoce Aerospace Technology Co., Ltd!

Tianjin GuoCe Aerospace Technology Co., Ltd.

Inspection Industry Solution Provider

We are more professional.

Testequipment

Telephone:18322499311

Mobile phone:18920407795

Fax:022-22145487

E-mail:10533646999@qq.com

Address:Chagugang Town, Wuqing District, Tianjin

Walk in salt spray laboratory

  • Hot Line:18322499311

Highlight-color: rgba(255, 255, 255, 0); margin: 0px; color: rgb(102, 102, 102); font-family: "microsoft yahei", 宋體, Arial; font-size: 14px;">


 


One Product Usage:


The walk-in salt spray test room is an upgraded salt spray test room based on the ordinary traditional salt spray test box. It can not only perform ordinary salt spray corrosion tests, but also high temperature and high humidity, low temperature and low humidity, and drying tests. At the same time, all tests are conducted under salt spray corrosion conditions. It has a large volume and can accommodate large metal components. The walk-in salt spray test room allows personnel to enter and exit, making it convenient for placing and removing specimens.


II Box structure:


The outer box is made of A3 steel plate sprayed with plastic or imported stainless steel material, while the inner box is made of highly corrosion-resistant plastic material or high-grade mirror stainless steel, which can effectively resist corrosion for up to ten years;


The water circulation device adopts a portable water storage tank, which can meet the long-term salt spray corrosion test;


The experimental chamber door is equipped with a large visible window, and the insulation material is made of polyurethane rigid foam;


The system control operation buttons and other components are installed in prominent and easy to maintain positions, and the door handle is a container opening door lock for the convenience of test personnel to open the door;


The tower spray system is equipped with a salt liquid filtration system, without a crystallization nozzle, the salt spray is evenly distributed, and the sedimentation is freely adjusted.


III Test details: (Composite salt spray test)


1. Neutral test


Laboratory: 35 ℃± 2 ℃ Saturated air tank: 47 ℃± 2 ℃


2. Acid test


Laboratory: 50 ℃± 2 ℃ Saturated air tank: 63 ℃± 2 ℃


3. Moisture test


Laboratory: 40 ℃± 2 ℃, relative humidity 90%~95% R.H (equipment can meet the set temperature up to 60 ℃, relative humidity up to 98% R.H)


Heating is achieved through an inner tank style heating and humidification method, which heats up quickly and distributes temperature and humidity evenly


Saturated air tank adopts automatic water replenishing. After continuous spray for about 20 hours, water replenishing will begin. After 1-2 minutes, water replenishing will end and spray will continue


4. Drying test


Drying treatment is carried out using a blower fan, and the specimen is dried for 24 hours under standard atmospheric conditions of temperature (15 ℃~35 ℃) and relative humidity not exceeding 50%


4 Control system:


Equipped with a Korean imported "ternary" temperature and humidity touch screen, a 7-inch high-definition true color LCD touch screen, and a controller operation interface with Chinese and English options for selection. The real-time operation curve chart can be displayed on the screen, with a capacity of 1000 program segments, each segment can cycle 999 steps, and the maximum time setting for each segment is 99 hours and 59 minutes. There are 10 sets of program linking functions;


The controller panel comes standard with a 10M/100M Ethernet interface, which automatically obtains IP addresses for remote control. It supports real-time monitoring, historical curve playback, program editing, FTP upload and download, historical fault viewing, remote setting/program control, and other functions.


Five Meet the standards:


The composite salt spray test chamber meets the requirements of coating quality for appearance parts of TSH1555G and TSH3131G, ASTMB-117, GB/T2423.17-2008, GB/T 2423.18-2000 "Basic Code for Electrical and Electronic Products - Test Ka", salt spray test method, and GB/T10125-1997 GB/T10587-2006、GB10593.2-1990、GB/T1765-1979、GB/T1771-2007、GB/T12967.3-2008、GB/T5170.8-2008 And equivalent IEC, MIL, DIN and other related standards.


VI Technical parameters:


 


Model (CM)


GC-YW-010


GC-YW-020


GC-YW-030


GC-YW-040


GC-YW-050


Studio size


85*130*60


90*200*60


95*250*150


100*270*170


110*290*180


External dimensions


115*210*155


125*280*160


135*290*170


135*300*180


145*310*190


Power


5.0(kw)


6.5(kw)


8.0(kw)


10(kw)


13(kw)


nature

can

finger

mark


temperature range 


Room temperature+5 ℃~55 ℃


Temperature uniformity


±2℃


Temperature fluctuation


±0.5℃


Salt spray deposition amount


1~2ml/80cm2.h


Spray mode


Choose between continuous and periodic options


Test timing


1~9999 (H, M, S) adjustable


temperature

wet

degree

control

system


temperature control


Korean ternary TEMI880 touch screen


Temperature sensor


Platinum resistance PT100Ω/MV


heating system 


Fully independent system, nickel chromium alloy electric heating heater


Spray system


Tower spray device with non crystallization nozzle (finer and evenly distributed particles)


Spray time


1~9999 (H, M, S) and adjustable cycle


Salt collection


Equipped with standard funnel and standard measuring cylinder


Salt liquid filtration


Install a water quality filter at the suction tube (to prevent nozzle blockage and terminate the test)


Preheating of salt solution


The temperature of the salt solution is balanced with the temperature inside the box (so that the low temperature of the salt solution does not affect the test temperature)


send

use

material

material


Cabinet material 


A3 steel plate spray coating or imported stainless steel material


Inner box material


Imported corrosion-resistant, high-strength P V. C/Imported corrosion resistance, high strength, aging resistance and high temperature resistance P P plate or high-grade mirror stainless steel plate


Other attachments


All are made of stainless steel, copper, or materials that are corrosion-resistant and heat-resistant


Standard configuration


1 round rod, 1 V-shaped sample holder, 2 nozzles, 2 funnels, and 2 sets of measuring cylinders


safeguard


Leakage, short circuit, overheating, water shortage, end of test, overcurrent protection


supply voltage


AC380V±10% 50Hz


Environmental temperature for use


5℃~+30℃ ≤85%R. H


seal up


Adopting a water sealed structure, there is no salt spray overflow


Note: This test chamber complies with GB2423.17-2008, as well as equivalent IEC, MIL, DIN, ASTM, etc

This test chamber can be extended to a humid and hot salt spray chamber according to relevant standards, with a humidity range of 20-98%

Can conduct separate wet heat test or salt spray test


With the advancement of technology, the above technical information is subject to change without prior notice!


   



Recommended products